Our development depends on the advanced equipment ,excellent talents and continuously strengthened technology forces for Ph Contror Dosing Pump , Ph Ec Monitor , Inline Electromagnetic Flow Transmitter , Currently, we are wanting ahead to even bigger cooperation with abroad customers according to mutual positive aspects. Be sure to sense free to contact us for more details.
Discount Price 2 In 1 Ph And Orp Controller - Online Conductivity TDS Sensor – JIRS Detail:

Pls unpack and check that the sensor has been supplied undamaged and that it is the correct option as ordered. If you have any problems please contact your supplier.

Application
Widely used for industrial water, tap water, cooling water, pure water etc. Conductivity measurement.

Main technique specification

Name

Function

Cell Constant

0.05cm-1  ( v )

0.1cm-1  (  )

1.0cm-1  (  )

10.0cm-1 (  )

Electrode structure

Bipolar

Electrode material

ABS (   )

316L Stainless steel(  v )

Temperature Sensor

NTC 10K ( v )

Pt 1000  (  )

 Pt 100  (  ) 

Thread structure

½” NPT Thread

Working pressure

0~0.5MPa

Operating temperature

0~50℃

Cable length

Standard:5m or others(5-30m)

Dimensions drawing
Conductivity-meter5

Conductivity(TDS)/ Resistivity electrode dimensions 

Installation and maintenance
Installation: To ensure the real measurement result, data distortion caused by air bubble or dead water in conductance cell should be avoided. The installation should be performed strictly according to the following drawing:

Conductivity-meter6

Notes
1. The electrode should be installed at a lower location in the pipe where the flow speed is steady and air bubbles are seldom generated.
2. No matter the conductance cell is horizontally or vertically installed, it should be deeply inserted into the moving water.
3. The conductivity/Resistivity signal is weak electronic signal and its collecting cable should be separately installed.
When threading cable joint or connecting terminal board is used, to avoid wetting interference or breakdown of measurement unit circuit, they should not be connected to the same group of cable joint or terminal board with the power or control line.
4. When the measurement cable needs to be lengthened, it’s recommended to use the cable provided by the original manufacturer, and the joint should be subject to reliable damp-proofing insulation disposal. When a longer distance is involved, the length of the cable (<30m) should be agreed on before delivery, and if the length is over 30m, a transmitter should be used.

Electrode Maintenance
1. The electrode cell should not be soaked in strong acid or alkali liquid, and the platinum black coating should not be wiped or it will lead to electrode surface damage and the constant and response capability will be affected. The right way should be: when the electrode is dirty, soak it in 10% dilute hydrochloric acid for a short while, then rinse it with pure water to keep the surface clean.
2. The measurement cable is special cable and should not be changed at will or it will cause significant error.

Joint wire
white wire to Cell -INPUT
Yellow wire to Cell -OUPUT
Black wire-TEMP
Red wire-TEMP

JiShen Water treatment Co., Ltd.
Add: No.18, Xingong Road, High-Technology Area, Shijiazhuang,China
Tel: 0086-(0)311-8994 7497 Fax:(0)311-8886 2036
E-mail:info@watequipment.com
Website: www.watequipment.com


Product detail pictures:

Discount Price 2 In 1 Ph And Orp Controller - Online Conductivity TDS Sensor – JIRS detail pictures

Discount Price 2 In 1 Ph And Orp Controller - Online Conductivity TDS Sensor – JIRS detail pictures

Discount Price 2 In 1 Ph And Orp Controller - Online Conductivity TDS Sensor – JIRS detail pictures


Related Product Guide:

The incredibly abundant projects administration experiences and 1 to one provider model make the superior importance of small business communication and our easy understanding of your expectations for Discount Price 2 In 1 Ph And Orp Controller - Online Conductivity TDS Sensor – JIRS , The product will supply to all over the world, such as: Montpellier, Mauritius, Barcelona, Our company abides by the management idea of "keep innovation, pursue excellence". On the basis of assuring the advantages of existing products, we continuously strengthen and extend product development. Our company insists on innovation to promote the sustainable development of enterprise, and make us become the domestic high-quality suppliers.
  • With a positive attitude of "regard the market, regard the custom, regard the science", the company works actively to do research and development. Hope we have a future business relationships and achieving mutual success.
    5 Stars By Lynn from Gambia - 2018.09.29 17:23
    It can be said that this is a best producer we encountered in China in this industry, we feel lucky to work with so excellent manufacturer.
    5 Stars By Victor from Buenos Aires - 2018.07.26 16:51
    Write your message here and send it to us