Fast delivery Electrode - Online PH transmitter with sensor – JIRS Detail:
Code: BSQ-PH-2019
General introduction
EC/ER/pH/ORP-2019 series transmitter mainly used for the water quality monitor sensor signal output, by 4-20mA, RS485 or TTL, the sensor output weak signal is enlarged to transmit, can be convenient for PLC, configuration software and SCM to communicate by adopting standard communication protocol MODBUS protocol.
EC/ER/pH/ORP series transmitter is widely used in metallurgy manufacturing, power, light industry, textile, water treatment equipment, water pipe network and the scientific research and other industrial sectors.
Transmitter series introduction
Note: A module of the transmitter can only be equipped with one
sensor, EC/ER is a module, and PH / ORP is a module.
①: EC conductivity transmitter; range: 0 ~ 4000us / cm, 0-10/20/200mS
②: ER resistivity transmitter; range: 0 ~ 18.2MΩ
③: pH transmitter; range: 0 ~ 14.00 pH
④: ORP redox potential transmitter; range: -2000 ~ + 2000mV
Transmitter series technique specification
No. |
Range |
Matched sensor |
Accuracy |
Connection |
1 |
0.1~18.25MΩ 0.05~10.00uS) |
1:316L S.S. plug into 0.01 sensor 2:Quick install 0.02 sensor; |
2 %FS |
1/2″NPT 1/4”Quick install |
2 |
0.1~200.0uS |
316L S.S. plug-into 0.1 sensor |
|
1/2″NPT) |
3 |
0.5~2000uS(standard) |
ABS1.0 Pt. Black sensor(standard) 316L S.S. plug into 1.0 sensor |
1.5% FS |
1/2″NPT
|
4 |
2~4000uS |
ABS1.0 Pt. Black sensor(standard) 316L S.S. plug into 1.0 sensor |
1.5% FS |
1/2″NPT
|
5 |
0.5~10mS |
ABS1.0 Pt. Black sensor(standard) 316L S.S. plug into 1.0 sensor |
3 % FS |
1/2″NPT
|
6 |
0.5~20mS |
1:316LS.S. plug into10.0 (standard) 2:PTFE+Titanium alloy 10.0 sensor |
1.5 % FS |
1/2″NPT 3/4″NPT |
7 |
0.5~100mS |
1:316LS.S. plug into10.0 (standard) 2:PTFE+Titanium alloy 10.0 sensor |
2 % FS |
1/2″NPT 3/4″NPT |
8 |
0.5~200mS |
PTFE+Titanium alloy 10.0 sensor |
2 % FS |
3/4″NPT |
PH transmitter technique specification
Measuring range: 0~14.00,
Accuracy:2.0%(FS),
Stability: ±2×10-3FS/24h,
Sensor: online PH sensor,
Sensor threaded size: 3/4″ BSP,
Cable length: standard length is 5m,
Measured media temperature:0~50℃,
Temperature sensor: NTC 10K,
Temperature compensation: On basis of 25℃, AUTO or Manual temperature compensation,
Calibration mode: Three points calibration,
PH4.00,PH6.86,PH9.18,
Other common technique specification
Output current: 4~20mA, isolated / Optional:1-5V /2-10V;
Relay output: High/Low limit relay alarm output, Contact point current 24V/3A, 220V/2A ( Passive dry contact)
Power: DC12V-28V, 24V(Current <=0.1A),
Environmental conditions: Working temperature:0~50℃, Relative humidity: ≤ 85%RH,
Overall dimension: 122×72×45mm(L x W x H),
Installation mode: Cabinet rail installation
Product detail pictures:
Related Product Guide:
To meet the customers' over-expected pleasure , we have now our solid crew to supply our greatest all round assistance which includes marketing, sales, planning, production, top quality controlling, packing, warehousing and logistics for Fast delivery Electrode - Online PH transmitter with sensor – JIRS , The product will supply to all over the world, such as: Kuwait, Swaziland, Juventus, With the goal of "zero defect". To care for the environment, and social returns, care employee social responsibility as own duty. We welcome friends from all over the world to visit and guide us so that we can achieve the win-win goal together.
By Giselle from United States - 2017.08.15 12:36
Product quality is good, quality assurance system is complete, every link can inquire and solve the problem timely!
By Janice from Greece - 2017.08.16 13:39