PFDO-800 Fluorescentia Dissolvenda Oxygen Sensor Operatio Manual

Description:

Sensorem oxygeni dissolutum metitur oxygenium dissolutum per methodum fluorescens, et lux emissa caerulea in stratum phosphori irradiatur.Substantia fluorescens excitatur ad lucem rubram emittere, et intentio oxygenii inverse proportionalis est tempori quo substantia fluorescens ad terram redit.Hac methodo utendo ut metiaris oxygenium dissolutum, non producet consumptionem oxygenii, ita certius data stabilitate, certa operatione, nullo impedimento, ac simplicem institutionem ac calibrationem.


Product Detail

Product Tags

Caput 1 Product Specifications

Specifications Singula
Magnitudo Diameter 49.5mm* Longitudo 251.1mm
Pondus 1.4KG
Principalis Material SUS316L+PVC (Versio Vulgata), Titanium Alloy (Version Seawater)
O-ring: Fluoro-rubber
Cable: PVC
IMPERVIUS Rate IP68/NEMA6P
Mensuratio dolor 0-20mg/L(0-20ppm)
Temperature: 0-45℃
Indicium Consilium Consilium: ±3%
Temperature: ±0.5℃
Repono Temperature -15~65℃
Environment Temperature 0~45℃
Pressura dolor ≤0.3Mpa
Potestas Supple 12 VDC
Calibration Automatic aeris calibratiis, Sample calibratiis
Cable Longitudo Standard 10-Meter Cable, Max Longitudo: 100 Meters
Warantum Tempus I anno
Dimensio externaPFDO-800 Fluorescens Oxygen Sensor Dissolvenda Operatio Manual4

Mensam I dissolvi Oxygen sensorem Technical Specifications

Caput II Product Information
Sensorem oxygeni dissolutum metitur oxygenium dissolutum per methodum fluorescens, et lux emissa caerulea in stratum phosphori irradiatur.Substantia fluorescens excitatur ad lucem rubram emittere, et intentio oxygenii inverse proportionalis est tempori quo substantia fluorescens ad terram redit.Hac methodo utendo ut metiaris oxygenium dissolutum, non producet consumptionem oxygenii, ita certius data stabilitate, certa operatione, nullo impedimento, ac simplicem institutionem ac calibrationem.
Productum late in plantis purgamentorum, aquae plantae, stationis aquae, aquae superficies, agriculturae, industriae et aliorum agrorum est.Species sensoris dissoluti oxygenii ostenditur sicut figura 1 .

PFDO-800 Fluorescentia Dissolvenda Oxygen Sensor Operatio Manual5

Figura I Dissolvetur Oxygen Sensor Aspectus

1. mensurae Cover

2- Temperature Sensor

3- R1

4- felis

5. Tutela cap

 

Chapter 3 Installation
3.1 instruitur sensoriis
Certae institutionis gradus haec sunt:
a.Instrue 8 (platea conscendens) in clausura piscinae cum 1 (fibulae M8 U-figura) ad sensorem ascendentem;
b.Coniungere 9 ad 2 (DN32) fistulam PVC glutine, dilabi sensorem per fistulam Pcv, donec cochleis sensoriis in 9 (adaptor), et curationem IMPERVIUS facias;
c.Fige 2 (DN32 tube) super 8 (laminam ascendentem) ab 4 (DN42U-figura Fibulae).

PFDO-800 Fluorescentia Dissolvenda Oxygen Sensor Operatio Manual6

Figura II Schematic Diagram in institutione sensoris

1-M8U-figura Clamp(DN60 2- DN32 Pipe (extra diametrum 40mm)
3 Hexagon Pentium Screw M6 * CXX " 4-DN42U-figura Pipe Clip
5- M8 Gasket(8*16*1) 6- M8 Gasket(8*24*2)
7. M8 vere Shim 8- Adscendens Plate
IX-Adaptor (Stread to Rectus-per)

3,2 Connection of Sensor
Sensorem recte coniungi debet cum hac definitione nuclei filum:

Vide non. 1 2 3 4
Sensor Cable Brown nigreos caeruleum album
Signum +12VDC AGND RS485 A RS485 B

Caput IV Calibratio Sensoris
Solutum oxygeni sensorem calibratum ad officinam est, et si te ipsum calibrare debes, gradus infra sequere.
Gradus specifici sunt hoc modo;
Duplex click "06", et archa pops e dextra.Mutare valorem ad 16 et deprime "mitte".

PFDO-800 Fluorescentia Dissolvenda Oxygen Sensor Operatio Manual8

② sensorem sicca et in aerem mitte, data mensurata stabilis est, duplex click "06", valorem ad 19 et deprime "mitte".

PFDO-800 Fluorescentia Dissolvenda Oxygen Sensor Operatio Manual7

Caput V Protocollum Communicationis
Sensorem instructum cum MODBUS RS485 functione communicationis, libet referre ad hanc sectionem manualem 3.2 ad reprimendam communicationem wiring.Default rate baud 9600 est, mensa specifica MODBUS RTU in sequenti tabula ostenditur.

MODBUS-RTU
Baud rate 4800/9600/19200/38400
Data Bits 8 bit
Pari Moderare no
Bit nolite 1bit
Nomen actis mandare OratioLocus DataType Longitudo Legere / scribere Descriptio  
Precium dissolvit Oxygen 0 F (Float) 2 R (tantum legitur)   Precium dissolvit Oxygen
Dissolvi Oxygenis Concentration 2 F 2 R   Dissolvi Oxygenis Concentration
Temperature 4 F 2 R   Temperature
Clivo 6 F 2 W/R Range:0.5-1.5 Clivo
Deviatio Value 8 F 2 W/R Range:-20-20 Deviatio Value
Salinity 10 F 2 W/R   Salinity
Pressura atmosphaerica 12 F 2 W/R   Pressura atmosphaerica
Baud rate 16 F 2 R   Baud rate
Servus Oratio 18 F 2 R Range: 1-254 Servus Oratio
Responsum tempus Read 20 F 2 R   Responsum tempus Read
Modift Baud Rate 16 Signum 1 W   0-48001-96002-19200

3-38400

4-57600

Mutare Servus Oratio 17 Signum 1 W Range: 1-254  
Mutare Responsio Tempus 30 Signum 1 W 6-60s Mutare Responsio Tempus
Air Calibration I step 27 Signum 1 W 16
Gradus 2 27 Signum 1 W 19
Deleri debet si calibrare non vis post supplicium "Gradus 1".
Cancel 27 Signum 1 W 21
Munus Code R:03
Scribere VI ut reshaing notitia VI "
Scribere XVI sicut fluctuetur notitia

Caput VI Sustentacionem
Ad optimos exitus mensurae obtinendas, valde necessarium est ut sensorem regulariter conservet.Tutela maxime includit purgationem, inspiciendo damnum sensoris et calibrationem periodicam.
6.1 Sensor Purgatio
Commendatur ut sensorem certis intervallis purgari debere (plerumque 3 menses, secundum situm ambituum) accurate mensurae curare.
Utere aqua emundare exteriorem superficiem sensorem.Si adhuc strages est, cum panno humido molliter extergito.Ne ponas sensorem in sole vel prope radialem directo.In tota vita sensoris, si totus sol patefactio temporis ad unam horam pervenerit, erit causa fluorescentis pilei senescentis et errati, ac per hoc ducens ad lectionem pravam.

6,2 metus de Damno Sensor
Secundum speciem sensoris reprimendam si damnum est;si quod damnum inventum est, pete contactum post-venditionis servitium centrum conservationis tempore ad reponendam ne malfunction sensoris causatur per aquam ex cap laesis.

6.3 De conservatione Sensoris
A. Cum illo non uteris, obtegere cap tutela primigenii producti ad vitandum solis vel nuditatis directum.Ut a frigore sensorem tuetur, DO probe reponendum est in loco ubi glaciatur non erit.
B. Retine specillum purga ante diu conditum.Serva apparatum in arca naviculae vel vase plastico cum praesidio electrico incursu.Vitare illam manu vel aliis rebus duris attingere in casu cappae fluorescentis scalpendi.
C. Prohibetur ut pileum fluorescens directo solis vel nuditatis exponatur.

Substitutio mensurae Cap 6,4
Mensurae pilei sensoris reponi debet cum laesum est.Ut accuratio mensurae provideatur, suadetur ut quotannis eam mutet vel reponi necesse est cum pileum graviter laesum in inspectione repertum est.

Caput VII Post-venditio Service
Si quas quaestiones habes vel officium reparandi opus est, pete nobis ut sequentia.