DO-6800

online Dissolved Oxygen Temperature Transmitting Controller

Measuring Range

0.00 – 20.00 mg /L ,   0 – 200.0 ug /L, (optional)

0-60℃

Resolution

0.1 ug/L,   0. 01 mg/L,   0.1℃

Accuracy

ug/L: ±1.0%FS; mg/L:±0.5%FS, ±0.3℃

Display

Large Screen Multi-parameters LCD

Temp. Compensation

NTC 10K, 0.0℃-60.0 ℃

Automatic temperature compensation

Current output

isolation to protect, 4 ~ 20mA signal output

Control output

two sets of ON / OFF relay contacts, divided into the high

point, low alarm signal, optically isolated output. 

Contact capacity

10A/220V AC (Resistive Load)

Output load

load <500Ω (0-10mA), load <750Ω (4-20mA)

Power

AC 220V±10%, 50/60Hz

Working environment

Ambient Temp. 0-60℃, Relative Humidity ≤90%

Dimensions

96×96×115mm (H X W X D), 0.9kgs

Hole size

91×91mm H X W)

Installation Mode

Panel Mounted

Protection Grade

IP57


Product Detail

Product Tags

Description:

  DO-6800 online Dissolved Oxygen Temperature Transmitting Controller is a micro computerized precision instrument.   

 The large LCD display, English menu operation.

  Multi-parameter display at the same time: the value of dissolved oxygen, temperature, output current, alarm points, etc. Intuitive and easy to read, and alarm for range over run

  Screen displays alarm status and accompanied by a switch ON / OFF signal output

●  Automatic Temperature Compensation: Automatic 0℃ to 60℃

  Communication function (optional):  RS-485 communication interface (MODBUS protocol is partially compatible), 4 to 20 mA current output corresponding to the DO value can be set freely.

  Notepad: can store 50 measurements, the user can record measurement data.

  Hysteresis amount can be set freely to avoid frequent switching relay action

  Watchdog function: to ensure that the instrument will not crash  

  Core devices from famous foreign brands

  Restore the factory settings. Power-down protection> 10 years

 

Application:

Widely used in the detection and control of dissolved oxygen in various water treatments, boiler water deoxygenization, environmental monitoring, aquaculture and other fields.

Main technique Specification:

Parameters

DO-6800

online Dissolved Oxygen Temperature Transmitting Controller

Measuring Range

0.00 – 20.00 mg /L ,   0 – 200.0 ug /L, (optional)

0-60℃

Resolution

0.1 ug/L,   0. 01 mg/L,   0.1℃

Accuracy

ug/L: ±1.0%FS; mg/L:±0.5%FS, ±0.3℃

Display

Large Screen Multi-parameters LCD

Temp. Compensation

NTC 10K, 0.0℃-60.0 ℃

Automatic temperature compensation

Current output

isolation to protect, 4 ~ 20mA signal output

Control output

two sets of ON / OFF relay contacts, divided into the high

point, low alarm signal, optically isolated output.

 

Contact capacity

10A/220V AC (Resistive Load)

Output load

load <500Ω (0-10mA), load <750Ω (4-20mA)

Power

AC 220V±10%, 50/60Hz

Working environment

Ambient Temp. 0-60℃, Relative Humidity ≤90%

Dimensions

96×96×115mm (H X W X D), 0.9kgs

Hole size

91×91mm H X W)

Installation Mode

Panel Mounted

Protection Grade

IP57