Online Resistivity transmitter with sensor

Short Description:

No.
Range
Matched sensor
Accuracy
Connection
1
0.1~18.25MΩ
0.05~10.00uS)
1:316L S.S. plug into 0.01 sensor
2:Quick install 0.02 sensor;
2 %FS
1/2″NPT
1/4”Quick install
2
0.1~200.0uS
316L S.S. plug-into 0.1 sensor
 
1/2″NPT)
3
0.5~2000uS
(standard)
ABS1.0 Pt. Black sensor(standard)
316L S.S. plug into 1.0 sensor
1.5% FS
1/2″NPT
4
2~4000uS
ABS1.0 Pt. Black sensor(standard)
316L S.S. plug into 1.0 sensor
1.5% FS
1/2″NPT
5
0.5~10mS
ABS1.0 Pt. Black sensor(standard)
316L S.S. plug into 1.0 sensor
3 % FS
1/2″NPT
6
0.5~20mS
1:316LS.S. plug into10.0 (standard)
2:PTFE+Titanium alloy 10.0 sensor
1.5 % FS
1/2″NPT
3/4″NPT
7
0.5~100mS
1:316LS.S. plug into10.0 (standard)
2:PTFE+Titanium alloy 10.0 sensor
2 % FS
1/2″NPT
3/4″NPT
8
0.5~200mS
PTFE+Titanium alloy 10.0 sensor
2 % FS
3/4″NPT


Product Detail

Product Tags

Code: BSQ-RM-2019
General introduction
EC/ER/pH/ORP-2019 series transmitter mainly used for the water quality monitor sensor signal output, by 4-20mA, RS485 or TTL, the sensor output weak signal is enlarged to transmit, can be convenient for PLC, configuration software and SCM to communicate by adopting standard communication protocol MODBUS protocol.

EC/ER/pH/ORP series transmitter is widely used in metallurgy manufacturing, power, light industry, textile, water treatment equipment, water pipe network and the scientific research and other industrial sectors.

Transmitter series introduction
Note: A module of the transmitter can only be equipped with one
sensor, EC/ER is a module, and PH / ORP is a module.

①: EC conductivity transmitter; range: 0 ~ 4000us / cm, 0-10/20/200mS
②: ER resistivity transmitter; range: 0 ~ 18.2MΩ
③: pH transmitter; range: 0 ~ 14.00 pH
④: ORP redox potential transmitter; range: -2000 ~ + 2000mV

Transmitter series technique specification

No.

Range

Matched sensor

Accuracy

Connection

1

0.1~18.25MΩ

0.05~10.00uS)

1:316L S.S. plug into 0.01 sensor

2:Quick install 0.02 sensor;

2 %FS

1/2″NPT

1/4”Quick install

2

0.1~200.0uS

316L S.S. plug-into 0.1 sensor

 

1/2″NPT)

3

0.5~2000uS(standard)

ABS1.0 Pt. Black sensor(standard)

316L S.S. plug into 1.0 sensor

1.5% FS

1/2″NPT

 

4

2~4000uS

ABS1.0 Pt. Black sensor(standard)

316L S.S. plug into 1.0 sensor

1.5% FS

1/2″NPT

 

5

0.5~10mS

ABS1.0 Pt. Black sensor(standard)

316L S.S. plug into 1.0 sensor

3 % FS

1/2″NPT

 

6

0.5~20mS

1:316LS.S. plug into10.0 (standard)

2:PTFE+Titanium alloy 10.0 sensor

1.5 % FS

1/2″NPT

3/4″NPT

7

0.5~100mS

1:316LS.S. plug into10.0 (standard)

2:PTFE+Titanium alloy 10.0 sensor

2 % FS

1/2″NPT

3/4″NPT

8

0.5~200mS

PTFE+Titanium alloy 10.0 sensor

2 % FS

3/4″NPT

ER- Resistivity transmitter technique specification
Measuring range:0~18.25MΩ
Accuracy:2.0%(FS),
Stability: ±2×10-3FS/24h,
Sensor: 0.01/0.02cm-1 S.S. 316L sensor, Plug-in (Cell=0.01) or quick install (Cell=0.02)
Sensor threaded size: 1/2″NPT,
Cable length: Plug-in type standard length is 5m,
quick mounted type is 1.5m,
Measured media temperature:0~50℃,
Sensor working pressure: 0-0.5Mpa,
Temperature sensor: NTC 10K,
Temperature compensation: On basis of 25℃, AUTO temperature compensation

Other common technique specification
Output current: 4~20mA, isolated / Optional:1-5V /2-10V
Relay output: High/Low limit relay alarm output, Contact point current 24V/3A, 220V/2A ( Passive dry contact)
Power: DC12V-28V, 24V(Current <=0.1A)
Environmental conditions: Working temperature:0~50℃, Relative humidity: ≤ 85%RH
Overall dimension: 122×72×45mm(L x W x H)
Installation mode: Cabinet rail installation


  • Previous:
  • Next:

  • Write your message here and send it to us